Example 1. Insulin Like Growth Factor I binding to immobilized Insulin Like Growth Factor Binding Protein 3

Instrument: Reichert SR7000DC Dual Channel SPR Spectrometer
Surface: Planar dithiol mixed self-assembled monolayer presenting –COOH, Catalog Number 13206062
Temperature: 25° C
Autosampler: Reichert/Spark Endurance
Immobilized Species: 950 µRIU Insulin Like Growth Factor Binding Protein 3
Analyte: Insulin Like Growth Factor 1 (IGF1)
flow rate: 50 µL/min
buffer: PBST
3 min 20 sec injections, 3 min 30 sec dissociations

Results:

ka (M-1 s-1) kd (s-1) KD, nM Surface
6.4e5 0.021 33.4 Planar

Tech Bulletin (PDF 52.2 KB)